Blood Coagulation Analyzer BK-CA02 BK-CA04 (biobase-0024) Products
Name Blood Coagulation Analyzer BK-CA02 BK-CA04
Code biobase-0024
Type Standard
Brand Biobase
Category Clinical Analytical Instruments
subcategory Coagulation Analyzer
Unit pcs (pcs)
test In Stock Yes
Berminat hub kami: wa chat
Product details for Invoice
Introduction:
The blood coagulation analyzer adopts the principle of optical colorimetry for detection. After the reagent is mixed with the plasma, fibrinogen is converted into fibrin and coagulated, which leads to the change of the optical density of the test specimen, and the instrument detects the coagulation end point.
Product Details
Blood Coagulation Analyzer BK-CA02 BK-CA04

Application:

It is suitable for the detection of prothrombin time(PT), activated partial thromboplastin time(APTT), thrombin time(TT) and fibrinogen(FIB) in plasma. .

Widely used in blood laboratory, biochemical laboratory, hospital blood laboratory and so on.

 

Features:

1. Dual-channel independent detection, can analyze different items at the same time

2. 5-inch high-definition touch screen, easy to operate

3. Open system, good compatibility, Supports reagents of various specifications

4. 20,000 results storage

5. Equipped with multiple pre-temperature positions to improve detection efficiency

6. Built-in thermal printer

7. The pipette automatically triggers the measurement, and the measurement time can be accurate to tenths of a second

8. Less reagent consumption

 

Technical Parameters:

Model

BK-CA02

BK-CA04

Test Principle

Optical colorimetry

Test Items

PT, APTT, TT, FIB

Test Channel

2(can test different projects at the same time)

4(can test different projects at the same time)

Wavelength

470nm

Detection Time

Normal sample detection time is 20 to 40 seconds, and can be set arbitrarily from 20 to 99 seconds

Sample Incubation Position

12

24

Reagent Incubation Position

5

6

Detection Position Temperature

37.0±1.0°C

Incubation Position Temperature

37.0±1.0°C

Incubation Time

1~999s

Reaction Time

1~99s

Reagent Dosage

Minimum dosage 20μl

Sample Volume

20μl~40μl

Display

5-inch high-definition color touch screen

Storage

20,000 results storage, automatically save measurement data when power off

Communication Interface

RS232 serial interface, support LIS system

Printer

Built-in thermal printer

Power Supply

AC100~240V, 50/60Hz

Consumption

70VA

Ambient Temperature

10℃~30℃

Relative Humidity

80%

Atmospheric Pressure

86kPa~106kPa

External Size(W* D*H)

360*210*120mm

390*250*135mm

Net Weight

3.5kg

4.96kg

Package Size(W* D*H)

376*246*316mm

445*305*265mm

Gross Weight

5kg

7kg